Measurement of glucose in ferrets: accuracy of three human portable meters

Authors

  • Daniel Alejandro Gómez Pizano Departamento de Etología, Fauna Silvestre y Animales de Laboratorio, Facultad de Medicina Veterinaria y Zootecnia, Universidad Nacional Autónoma de México
  • Ángela Rodríguez Hernández Departamento de Etología, Fauna Silvestre y Animales de Laboratorio, Facultad de Medicina Veterinaria y Zootecnia, Universidad Nacional Autónoma de México
  • Adriana Margarita Ducoing Watty partamento de genética y bioestadística, Facultad de Medicina Veterinaria y Zootecnia, Universidad Nacional Autónoma de México
  • Delia Arlette Castillo Mata Departamento de Patología, Facultad de Medicina Veterinaria y Zootecnia, Universidad Nacional Autónoma de México
  • Ricardo Itzcóatl Maldonado Reséndiz Departamento de Etología, Fauna Silvestre y Animales de Laboratorio, Facultad de Medicina Veterinaria y Zootecnia, Universidad Nacional Autónoma de México

DOI:

https://doi.org/10.19230/jonnpr.3044

Keywords:

Glucose, point-of-care glucometers, glycemia, Mustela

Abstract

Aims. In ferrets, portable equipment could underestimate blood glucose values, this study evaluate the  accuracy of these meters.

Methods and Material. Blood samples were obtained from 11 ferrets to compare glycemia values obtained with 3 portable blood glucometers.

Results. The 3 point-of-care glucometers tested underestimated the glycaemia values when compared with the results obtained with the automated biochemical analyzer. Therefore, the results  suggest that none of the 3 portable blood glucometers are reliable for measuring blood glucose in ferrets,  also indicates an average underestimation in relation to the laboratory analyzer.

Conclusions. The clinicians should be cautious when interpreting glycemia values in ferrets obtained with these 3 point-of-care glucometers, and always corroborate the results with an automated biochemical analyzer.

 

Downloads

Download data is not yet available.

Author Biography

Daniel Alejandro Gómez Pizano, Departamento de Etología, Fauna Silvestre y Animales de Laboratorio, Facultad de Medicina Veterinaria y Zootecnia, Universidad Nacional Autónoma de México

Especialista en medicina y cirugía en modalidad fauna silvestre, profesor de asignatura facultad de medicina veterinaria y zootecnia UNAM, académico del hospital de especialidades en fauna silvestre UNAM, Jefe de área operativa de atención a peces del zoológico Benito Juarez Morelia.

References

Cohn LA, Mccaw DL, Tate DJ, Johnson JC. Assessment of five portable blood glucose meters, a point-of-care analyzer, and color test strips for measuring blood glucose concentration in dogs. J Am Vet Med Assoc 2000; 216: 198-202.

Wess G, Reusch C. Assessment of five portable blood glucose meters for use in cats. Am J Vet Res 2000; 61: 1587-1592.

Wess G, Reusch C. Evaluation of five portable blood glucose meters for use in dogs. Am J Vet Res 2000; 216: 203-209.

Domori A, Sunahara A, Tateno M, Miyama TS, Setoguchi A, Endo Y. The clinical utility of two human portable blood glucose meters in canine and feline practice. Vet Clin Pathol 2014; 43: 55-62.

Cohen TA, Nelson RW, Kass PH, Christopher MM, Feldman EC. Evaluation of six portable blood glucose meters for measuring blood glucose concentration in dogs. J Am Vet Med Assoc 2009; 235: 1-5.

Brito-Casillas Y, Figueirinhas P, Wiebe JC, López-Ríos L, Pérez-Barreto D, Melián C et al. ISO-Based Assessment of Accuracy and Precision of Glucose Meters in Dogs. J Vet Intern Med 2014; 28: 1405-13.

Petritz OA, Antinoff N, Chen S, Kass PH, Paul-murphy JR. Evaluation of portable blood glucose meters for measurement of blood glucose concentration in ferrets. J Am Vet Med Assoc 2013; 242: 350-354.

Lewington JH. Ferret handling, hospitalization and diagnostic techniques. En: Lewington JH, editor. Ferret Husbandry, Medicine and Surgery 2da ed, Philadelphia: ElServier Saunders; 2007, p. 151-165.

Siperstein LJ. Ferret hematology and related disorders. Vet Clin North Am Exot Anim Pract 2008; 11: 535-50, vii.

Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Int J Nurs Stud 2010; 47: 931-936.

Altman DG. Practical Statistics for Medical Research. 1st ed. London: Chapman and Hall/ CRC; 1990.

Summa NM, Eshar D, Lee-chow B, Larrat S, Brown DC. Comparison of a human portable glucometer and an automated chemistry analyzer for measurement of blood glucose concentration in pet ferrets. Can Vet J 2014; 55: 865-869.

Johnson BM, Fry MM, Flatland B, Kirk CA. Comparison of a human portable blood glucose meter, veterinary portable blood glucose meter, and automated chemistry analyzer for measurement of blood glucose concentrations in dogs. JAVMA 2009; 235: 1309-1313.

Selleri P, Girolamo N Di, Novari G. Performance of two portable meters and a benchtop analyzer for blood glucose concentration measurement in rabbits. JAVMA 2014; 245: 87-97.

Acierno MJ, Mitchell MA, Schuster PJ, Freeman D, Guzman DS, Jr TNT. Evaluation of the agreement among three handheld blood glucose meters and a laboratory blood analyzer for measurement of blood glucose concentration in Hispaniolan Amazon parrots (Amazona ventralis). AJVR 2009; 70.

Published